Lateral temperature spread of monopolar, bipolar and ultrasonic instruments for robot-assisted laparoscopic surgery

Hefermehl LJ, Largo RA, Hermanns T, Poyet C, Sulser T, Eberli D

Published August 2014 BJU International, 114(2), 245-252
DOI 10.1111/bju.12498 PMID 24127773

RRM Academy Synopsis

Monopolar Cautery Spreads Far More Lateral Heat Than Bipolar Tools

Monopolar cautery tools spread heat beyond 20 millimeters within four seconds at 60 watts, in a bench study on strips of cow muscle tissue. Bipolar, vessel-sealing, and ultrasonic tools stayed under 5 millimeters even at their longest settings. Researchers tested these instruments, which are commonly used in robotic and laparoscopic surgery, under controlled lab conditions.

Study at a glance, bench study: Monopolar cautery spread heat much farther than bipolar or ultrasonic tools
Study at a glance, bench study: Monopolar cautery spread heat much farther than bipolar or ultrasonic tools. Source: Hefermehl LJ et al., 2014, PMID 24127773.

Key Findings

  • At 60 watts for one second, monopolar instruments produced thermal spreads of 3.5 mm (Hot Shears) and 2.5 mm (Cautery Hook). Both exceeded 20 mm by four seconds.
  • Bipolar forceps produced smaller spreads of 2.1 mm (Maryland) and 2.2 mm (PreCise) at 60 watts for one second, rising to 3.6 mm and 2.9 mm at two seconds.
  • Vessel-sealing devices (PK forceps, LigaSure) and ultrasonic shears reached thermal spreads of up to 3.9 mm (PK forceps) and 2.9 mm (Harmonic shears) at their longest tested settings.
  • Adding a second instrument as a heat sink next to the cautery device reduced spread for most tools, for example from 2.5 mm to 1.4 mm for the Cautery Hook.
  • Staining for cell damage at 60 W showed intact tissue at 2.4 mm (Hot Shears) and 2.2 mm (Cautery Hook); bipolar forceps reached only 1.3 mm (Maryland) and 1.6 mm (PreCise).

Interpretation

This bench study tested seven cautery and vessel-sealing tools on strips of fresh cow muscle at one Swiss academic center. Each thermal measurement was repeated six times and analyzed with infrared imaging. A separate cell-damage stain checked tissue injury at the protein level. The authors note that this in vitro method does not fully reflect a real operation, though it has closely matched living-tissue results in earlier work. The findings support larger studies of these thermal differences in live surgical settings.

RRM Context

Restorative reproductive medicine protects nerves and the fallopian tubes during pelvic surgery. Surgeons who treat endometriosis work close to these parts. This paper comes from urology. The mechanism it describes, heat spreading sideways from cautery tools, is relevant to laparoscopic and robotic procedures that work near delicate pelvic tissue more broadly.

Abstract

Objective

To assess critical heat spread of cautery instruments used in robot-assisted laparoscopic (RAL) surgery.

Materials and Methods

Thermal spread along bovine musculofascial tissues was examined by infrared camera, histology and enzyme assay. Currently used monopolar, bipolar and ultrasonic laparoscopic instruments were investigated at various power settings and application times. The efficacy of using an additional Maryland clamp as a heat sink was evaluated. A temperature of 45 °C was considered the threshold temperature for possible nerve damage.

Results

Monopolar instruments exhibited a mean (sem) critical thermal spread of 3.5 (2.3) mm when applied at 60 W for 1 s. After 2 s, the spread was >20 mm. For adjustable bipolar instruments the mean (sem) critical thermal spread was 2.2 (0.6) mm at 60 W and 1 s, and 3.6 (1.3) mm at 2 s. The PK and LigaSure forceps had mean (sem) critical thermal spreads of 3.9 (0.8) and 2.8 (0.6) mm respectively, whereas the ultrasonic instrument reached 2.9 (0.8) mm. Application of an additional Maryland clamp as a heat sink, significantly reduced the thermal spread. Histomorphometric analyses and enzyme assay supported these findings.

Conclusions

All coagulation devices used in RAL surgery have distinct thermal spreads depending on power setting and application time. Cautery may be of concern due to lateral temperature spread, causing potential damage to sensitive structures including nerves. Our results provide surgeons with a resource for educated decision-making when using coagulation devices during robotic procedures.